- Topic ID: id_16157848
- Version: 1.0
- Date: Jul 7, 2018 4:34:48 PM
Axial Encoder Assembly Replacement
Prerequisites
Overview
This replacement procedure defines the replacement process for the axial encoder assembly.
Procedure
- Remove gantry covers as required.
Refer to
- Tilt gantry back to 30 degrees.
- notice
- Turn OFF all 3 gantry service switches (Axial Drive, HVDC, 120VAC).
- Disconnect the Encoder DB 9 pin connect from the gantry harness.
- Carefully cut tie-wraps as necessary.
- Using an 8 mm hex key, remove the shoulder screw on which the
encoder assembly pivots.note:
Do not remove grease from the screw shoulder. Some grease is necessary for free movement of the assembly. If necessary wipe some grease from the inside of the old assembly and re-apply to the shoulder bolt prior to installing the new assembly.
- Install in reverse order.note:
Make sure the shoulder bolt has some grease on the shoulder of the bolt as noted in the prior step.
note:Apply a minimal amount of Loctite 242 to the tip of the shoulder bolt threads. This is to insure that the Loctite does NOT wick out onto the shoulder of the bolt and mix with the grease.
- Perform “Resetting the C-Pulse” (located in folder).
![]() |
Finalization
- Perform Axial Encoder Check, located in Axial Motion Checks.
- Perform Axial Control Functional, located in Axial Motion Checks.
- Perform System Scanning Test (located in folder).