Over-current Protection |
OC1 |
Over-current occurred during acceleration. |
The inverter output current momentarily exceeded
the over-current level. |
The inverter output terminals were short-circuited. |
Remove the wires connected to the inverter output
terminals (U, V, and W) and measure the interphase resistance. Check
if the resistance is too low. |
|
OC2 |
Over-current occurred during deceleration |
|
OC3 |
Over-current occurred when running at a constant speed |
|
Over-voltage Protection |
OU1 |
Over-voltage occurs during the acceleration |
The DC link circuit voltage was over the detectin
level of over-voltage. |
The power supply voltage was over the range of
the inverter's specifications. |
Measure the input voltage. ⇒ Decrease
the voltage to within that of the specifications.
|
|
OU2 |
Over-voltage occurs during the deceleration |
|
OU3 |
Over-voltage ccurs during the running at constant speed |
|
Under-voltage Protection |
LU |
The DC link circuit voltage
was below the under-voltage detection level |
The power supply voltage did not reach the range of the inverter's
specifications. |
Measure the input voltage. ⇒ Increase the voltage
to within that of the specifications.
|
|
Input Phase Loss Protection |
L in |
Input phase loss
occurred, or interphase voltage unbalance rate was large. |
Main circuit power input wires broken. |
Measure the input voltage. ⇒ Repair or replace the
wires.
|
| The terminal screws for the main circuit power input of the
inverter were not tight enough. |
Check if the screws on the inverter input terminals have become
loose. ⇒ Tighten the terminal screws to the recommended
torque.
|
|
Output Phase Loss Protection |
OPL |
Output phase loss occurred. |
The terminal screws for inverter output were not tight enough. |
Check if any scew on teh inverter output terminals has become
loose. ⇒ Tighten the terminal screws to the recommended
torque.
|
|
Overheat Protection for Heat Sink |
OH1 |
Temperature around
heat sink rose. |
Temperature around the inverter exceeded that of inverter specifications. |
If the temperature around the inverter is normal, replace the
inverter. |
| Air vent is blocked. |
Check if the heat sink is not clogged. ⇒ Clean the
heat sink.
|
|
Overload Protection |
OLU |
Temperature inside
inverter rose abnormally. |
Temperature around the inverter exceeded that of inverter specifications. |
If the temperature around the inverter is normal, replace the
inverter. |
| Air vent is blocked. |
Check if the heat sink is not clogged. ⇒ Clean the
heat sink.
|
note:
An under bar ( _ _
_ _ ) will be displayed when an under-voltage condition is detected
and a run command is present while the setting of F14 (Restart mode
after instantaneous power failure (function selection )) is not ”0”.
|