• Topic ID: id_11039102
  • Version: 2.0
  • Date: Jan 30, 2019 9:43:04 PM

1X Image Series Outline

1 GE Performance Phantom

  • MTF

    • Small Spot (< 200mA)

      • Standard Algo.

      • 3.8 lp/cm @ 50% Modulation

      • 6.5 lp/cm @ 10% Modulation

    • Large Spot (> 200mA)

      • Standard Algo.

      • 3.8 lp/cm @ 50% Modulation

      • 6.2 lp/cm @ 10% Modulation

  • MTF

    • Small Spot (< 200mA)

      • Edge Algo.

      • 8.5 lp/cm @ 50% Modulation

      • 13.0 lp/cm @ 10% Modulation

    • Large Spot (> 200mA)

      • Edge Algo.

      • 7.9 lp/cm @ 50% Modulation

      • 12.1 lp/cm @ 10% Modulation

2 GE QA Phantom

MTF

  • Standard Algo.

  • MTF 0.65 - 1.0

  • Contrast Scale 110 to 130

3 5” Water Phantom

Center Spot