- Topic ID: id_11039102
- Version: 2.0
- Date: Jan 30, 2019 9:43:04 PM
1X Image Series Outline
1 GE Performance Phantom
-
MTF
-
Small Spot (< 200mA)
-
Standard Algo.
-
3.8 lp/cm @ 50% Modulation
-
6.5 lp/cm @ 10% Modulation
-
-
Large Spot (> 200mA)
-
Standard Algo.
-
3.8 lp/cm @ 50% Modulation
-
6.2 lp/cm @ 10% Modulation
-
-
-
MTF
-
Small Spot (< 200mA)
-
Edge Algo.
-
8.5 lp/cm @ 50% Modulation
-
13.0 lp/cm @ 10% Modulation
-
-
Large Spot (> 200mA)
-
Edge Algo.
-
7.9 lp/cm @ 50% Modulation
-
12.1 lp/cm @ 10% Modulation
-
-
2 GE QA Phantom
MTF
-
Standard Algo.
-
MTF 0.65 - 1.0
-
Contrast Scale 110 to 130
3 5” Water Phantom
Center Spot