Gantry Tilt Verification

Prerequisites

Overview

This test is a verification of the relationship of the gantry tilt to the table level in the Y plane.

Procedure

  1. From the Install folder of the service manual, select the Installation ManualBook 2 Electrical Calibration.
  2. Perform the Tilt Alignment Tool Procedure From the Image Quality chapter.
  3. Remove gantry base and table base covers as applicable for adjustments required by the alignment procedure.

    Refer to Parts Replacement > Gantry > Enclosure > (Cover Removal Procedures).

    Refer to Parts Replacement > Table > (Cover Removal Procedures).

Finalization

  1. Replace Table and Gantry covers removed in prior steps.

    Refer to Parts Replacement > Gantry > Enclosure > (Cover Removal Procedures).

    Refer to Parts Replacement > Table > (Cover Removal Procedures).

  2. Finalization is complete.