- Topic ID: id_15460219
- Version: 5.0
- Date: Jan 20, 2020 8:35:57 PM
Gantry Rotation Safety Check
Prerequisites
Overview
Use DDC to Rotate the Gantry for 10 minutes, at the fastest allowed gantry speed.
Procedure
- Make sure the Axial Drive Enable switch (S2) and HVDC swich (S3) on the service switch panel are off.
- Manually rotate the gantry through a full rotation listening
for an contact between stationary and rotating parts
-
Listen for an contact between stationary and rotating parts. Correct any issues found before continuing.
-
Watch and listen for any loose parts/cables and tighten them before continuing.
-
- Turn on the HVDC and Axial Drive service switches.
- notice
- Start the Axial Functional diagnostic.
- Open the Common Service Desktop
- Select the Diagnostics Tab
- Select the Diagnostic Data Collection menu item to start the diagnostic.
- notice
- Select a rotation speed of 2 seconds and run the test. This
step should be run for a minimum of 30 seconds.
If using DDC/ Rotating X-Ray off, make sure Rotor Overrides is on (or else it will brake a LB rotor)
Listen for an contact between stationary and rotating parts. Correct any issues found before continuing.
note:Be ready to click on the Stop button on the menu immediately if you hear any problems. (This stops the gantry faster than the red emergency stop (e-stop) button)
- Repeat Step 5 for rotation speeds of 1.0 and 0.5 seconds.
- Wait 3 minutes after the gantry stops rotating before continuing to allow for axial drive cooling.
- Rotate the gantry at 0.35 seconds/revolution for 60 seconds
(to give the gantry time to get up to speed).
Listen for an contact between stationary and rotating parts. Correct any issues found before continuing.
note:Be ready to click on the Stop button on the menu immediately if you hear any problems. (This stops the gantry faster than the e-off button)
- If no issues arise, gantry rotation capability is safe for use.
![]() |
![]() |
Finalization
No finalization steps.