• Topic ID: id_15460102
  • Version: 3.0
  • Date: Jun 15, 2020 11:00:24 PM

GOC6.6 VCT Host Computer (Z820) NIC Card Replacement

Prerequisites

Overview

This procedure shall be followed when replacing the NIC Card in the GOC6.6 VCT host computer (Z820).

1 Z820 Computer Removal

Procedure

  1. Shutdown system. Select one of the following methods to Power OFF the Console:
    • If Applications are up, click on the [Shut Down] button on desktop display and select Shutdown.

    • If Applications are down, open a Terminal Window. Type: halt , then press ENTER.

    • When halt command has finished, power Off the console at the front panel switch.

  2. Apply LOTO. See Equipment Service - Lockout-Tagout-PPE procedure.
  3. Remove Keyboard Tabletop, Front and Rear covers from GOC6.6 console.

    Refer to Replacement → Console (GOC6.6) → Console Cover Removal and Installation.

  4. Remove the Z820 computer from the GOC6.6 console chassis.

    Refer to Replacement → Console (GOC6.6) → GOC6.6 VCT Host Computer (Z820) Replacement

2 NIC Card Replacement

Procedure

  1. Open the Z820 computer side access panel.

    Figure 1. Z820 Side Access Panel Removal

  2. Remove the Expansion Card Support.

    Figure 2. Z820 Airflow Guide and Expansion Card Support

  3. Remove the following components from the Z820 computer:
    1. FDIP Card (Slot 1)
    2. Quadro 2000 Graphic Card (Slot 2)
    3. RAID Controller (Slot 4)

      Figure 3. Z820 Component Location

  4. Replace the existing Ethernet NIC card with new one.
  5. Reinstall the removed PCI cards to the Z820 computer in the following order:
    1. RAID Controller (Slot 4)
    2. Quadro 2000 Graphic Card (Slot 2)
    3. FDIP Card (Slot 1)
  6. Install the Expansion Card Support and close the Side Access Panel.

3 Restore the Console

Procedure

  1. Fully slide the Z820 computer back into the console chassis, until mounting brackets are flush against the console chassis.
  2. Remove and store the Service Platform.
  3. Install and torque the four (4) 3mm socket cap screws that mount the Z820 computer in the GOC6.6 console.
    note:

    Torque each the for (4) 3mm socket cap mounting screws to:

  4. Reconnect all cables removed earlier to the Z820 computer.
  5. Remove LOTO on console.

4 Finalization

Procedure

  1. Install console covers and Keyboard Tabletop assembly. Refer to Replacement → Console (GOC6.6) → Console Cover Removal and Installation Procedure.
  2. Perform the Functional Checks → System Scanning Test instructions from the procedure list.
  3. Confirm Z820 computer does not display POST errors. Confirm computer functionality during Finalization section of this procedure.
    note:

    New Software Option Licenses required !