Foot Switch End Assy Replacement
Prerequisites
Overview
Procedure
- Raise the Table to maximum height.
- Move the Cradle and IMS to OUT limit position.
- Remove power from Table by turning off 120VAC, Axial Drive and HVDC switches on Service Switch Panel.
- Remove 4 foot pads from the foot switch assembly.
- Remove a foot switch cover by unscrewing 3 screws.
- Unscrew 4 screws holding the Foot Switch End Assy, and disconnect
the cable connector, then remove the Foot Switch End Assy from the
foot switch assembly.
Figure 1. Foot Switch End Assy Removal

- Install the new Foot Switch End Assy by referring to Step 6 through Step 4.
Finalization
- Power up the Table from the Service Switch Panel.
- Verify that the foot switch function is operating normally.