Collimator Calibration

Prerequisites

Overview

This document provides necessary steps to detailed calibrate the system.

Procedure

  1. Move table to longitudinal home position.
  2. Click SCANNER UTILITIES.
  3. Click COLLIMATOR CALIBRATION.
  4. Click OK in Attention window.
  5. Press START on SCIM when it flashes.
  6. Click OK in Attention window.
  7. Click QUIT in Scanner Utilities window.
    note:

    If this procedure has been done as part of Tube Replacement procedure, return to Tube replacement procedure for remaining steps

Finalization

  1. Detailed Calibration
  2. Fast Calibration
  3. Save System State