Center Phantom
Prerequisites
1 Prepare the QA Phantom
note:
The QA phantom is shipped water-filled.
Procedure
- Locate the multi-language sticker packet in the QA phantom shipping box.
- Attach the sticker with the customer's language to the face of the phantom hanger bracket.
- Check for bubbles.
- Purge and refill if necessary.
2 Center Phantom
Procedure
- Locate the QA phantom and mount it and the phantom holder on the table.
- Use the adjustment knobs on the phantom holder to level the phantom front-to-back and sideto- side with a bubble level.
- Select SCANNER UTILITIES on the left monitor.
- Select CENTER PHANTOM.
- Follow the on-screen procedures.note:
The phantom center spec is ± 0.5mm.
- Select QUIT when the phantom is within
specification.
Level the phantom both front-to-back and side-to-side. (Use a 6" level.)
Finalization
No finalization steps.