Center Phantom

Prerequisites

1 Prepare the QA Phantom

note:

The QA phantom is shipped water-filled.

Procedure

  1. Locate the multi-language sticker packet in the QA phantom shipping box.
  2. Attach the sticker with the customer's language to the face of the phantom hanger bracket.
  3. Check for bubbles.
  4. Purge and refill if necessary.

2 Center Phantom

Procedure

  1. Locate the QA phantom and mount it and the phantom holder on the table.
  2. Use the adjustment knobs on the phantom holder to level the phantom front-to-back and sideto- side with a bubble level.
  3. Select SCANNER UTILITIES on the left monitor.
  4. Select CENTER PHANTOM.
  5. Follow the on-screen procedures.
    note:

    The phantom center spec is ± 0.5mm.

  6. Select QUIT when the phantom is within specification.

    Level the phantom both front-to-back and side-to-side. (Use a 6" level.)

Finalization

No finalization steps.