• Topic ID: id_15460290
  • Version: 3.0
  • Date: Jun 15, 2020 10:56:49 PM

(13HW31.8) Service Pack 2.1 Installation Procedure

Prerequisites

Overview

This procedure describes 13HW31.8 Service Pack 2.1 installation procedure for GOC6.6 (Z820).

The Service Pack 2.1 software CDROM contains Service Pack 1.0, Service Pack 2.0 and Service Pack 2.1 patches.

The Service Pack 2.1 is NOT applicable for GOC5, 6, 6.5 and GOC6.6(Z840).

1 Preparation for Installation

Procedure

  1. Check the current software version by the following command.

    {ctuser@hostname} swhwinfo Enter

    Example: 13HW31.8_SP2-0-1.V40_H_HS64_G_GTL

    If the output contains SP2-1-x, SP2.1 is already installed. No need to install it.

    If the output contains SP2-0-x, SP2.0 is installed. Go to section 4.2 Service Pack 2.1 Installation (on SP2-0-x software).

    If the output contains SP1-0-x, SP1.0 is installed. Go to section 4.3 Service Pack 2.1 Installation (on SP1-0-x software).

    If the output is just 13HW31.8, no SP is installed. Go to section 4.4 Service Pack 2.1 Installation (on no SP software).

    If the output is not equal to 13HW31.8, system software version is different, do not execute the procedure.

  2. Check Exam Split Type.

    Type swokinstall -p

    If Exam Split option is displayed, execute the following

    Type ls –l ~ctuser/ves/.hesModeEnter

    If there is .hesMode file, it’s HES mode.

    If not, It’s VES mode.

    This will be used at reinstall option.

2 Service Pack 2.1 Installation (on SP2-0-x software)

Procedure

    Notice Image
    • notice
  1. Execute “Save Runtime Parameters” from Generator tool – JEDI in Diagnostic tab of CSD menu.
  2. Open a Unix shell.
  3. If Application software is up, perform the following command in the Unix shell.

    {ctuser@hostname} cleanMon

  4. In the Unix Shell, become root:
    1. {ctuser@hostname} su -
    2. Password: <password>
  5. Insert the Service Pack CD into DVD drive in the host.
  6. Perform the following steps to install the Service Pack CD:
    1. [root@hostname]# patch_install -c
    2. The following messages is displayed.


      These patches from cdrom will be installed:
      13HW31.8_GEHCServicePack2.1
      ------------------------------------------
      I will install update 13HW31.8_GEHCServicePack2.1 from CDROM, is this ok ? [y/n]

                        

    3. Input y to install Service Pack 2.1.

      The following message appears after completion of SP2.1 installation.


      #################################################################
      # Installation successful.                                      #
      #                                                               #
      # - Execute reconfig in order to set Bearing type (Ball/Liquid) #
      #   after SP2.1 installation completion.                        #
      # - If NTP server is used, confirm and accept the network       #
      #   setting by Reconfig.                                        #
      #################################################################

                        

    4. Confirm the “Update 13HW31.8_GEHCServciePack2.1 has been installed successfully” message.
  7. Check Software version:
    1. {root@hostname]# swhwinfo Enter
    2. System should reply with a response as the following:

      13HW31.8_SP2-1-X. <hw info (ex. V40_H_H64_G_GTL)>

  8. Go to reconfig. menu and set Bearing type to match the currently installed X-ray Tube. Refer to System Configuration (12HW14.6 SP2.1 or later).

    If NTP server is used, confirm and accept the network setting by Reconfig.

    Reboot is executed during reconfig when bearing type was updated. In this case, skip step 9 “Reboot the system”.

  9. Remove the Service Pack CD from Host DVD drive.
  10. Reboot the system:

    {root@hostname]# reboot

  11. Execute “Flash Download” in Utilities tab of CSD menu.
  12. Type “swhwinfo” in command window and confirm that the tube type is same as bearing type in reconfig.

    Example of strings: 13HW31.8_SP2-1-3.V40_NV_V64_G_GTM in case of VCT Liquid Bearing.

    • Ball Bearing : H

    • Liquid Bearing: NV

  13. Execute “Restore Runtime Parameters” from Generator tool – JEDI in Diagnostic tab of CSD menu.
  14. Execute “Generator Filament Calibration” in Calibration tab of CSD menu.
  15. Perform sanity scan with the following technique and confirm they completed successfully.

    Scout, Axial and Helical.

3 Service Pack 2.1 Installation (on SP1-0-x software)

Procedure

    Notice Image
    • notice
  1. Execute “Save Runtime Parameters” from Generator tool – JEDI in Diagnostic tab of CSD menu.
  2. Execute “Save System State” in Utilities tab of CSD menu.
  3. Uninstall the following options (if installed) using the Install Options menu in the Configuration Tab in Common Service Desktop.
    • AdvVesselAnalysis

    • AutoBone

    • CardEP

    • CardIQPlus

    • CTColonoPro

    • CTPerfusion4

    • CTPerfusion4Neuro

    • VolumeViewer

  4. Open a Unix shell.
  5. If Application software is up, perform the following command in the Unix Shell.

    {ctuser@hostname} cleanMon

  6. In the Unix Shell, become root:
    1. {ctuser@hostname} su -
    2. Password: <password>
  7. Insert the Service Pack CD into DVD drive in the host.
  8. Perform the following steps to install the Service Pack CD:
    1. [root@hostname]# patch_install -c
    2. The following message is displayed.


      These patches from cdrom will be installed:
      13HW31.8_GEHCServicePack2.0
      ------------------------------------------
      I will install update 13HW31.8_GEHCServicePack2.0 from CDROM, is this ok ? [y/n]

                        

    3. Input y to install Service Pack 2.0.

      The following message appears after completion of SP2.0 installation. Ignore it at this timing.


      #################################################################
      # Installation successful.                                      #
      #                                                               #
      # - Execute reconfig in order to set Bearing type (Ball/Liquid) #
      #   after SP2.0 installation completion.                        #
      # - If NTP server is used, confirm and accept the network       #
      #   setting by Reconfig.                                        #
      # - Reinstall the software options and Voxtool protocols by     #
      #   Restore State.                                              #
      #################################################################

                        

    4. Confirm the “Update 13HW31.8_GEHCServciePack2.0 has been installed successfully” message.
    5. The following message is displayed.


      These patches from cdrom will be installed:
      13HW31.8_GEHCServicePack2.1
      ------------------------------------------
      I will install update 13HW31.8_GEHCServicePack2.1 from CDROM, is this ok ? [y/n]

                        

    6. Input y to install Service Pack 2.1.

      The following message appears after completion of SP2.4 installation.


      #################################################################
      # Installation successful.                                      #
      #                                                               #
      # - Execute reconfig in order to set Bearing type (Ball/Liquid) #
      #   after SP2.1 installation completion.                        #
      # - If NTP server is used, confirm and accept the network       #
      #   setting by Reconfig.                                        #
      #################################################################

                        

    7. Confirm the “Update 12HW14.6_GEHCServciePack2.1 has been installed successfully” message.
  9. Check Software Version:
    1. Type: [root@hostname]# swhwinfo ENTER
    2. System should reply with a response similar to the following:

      13HW31.8_SP2-1-X.<hw info (ex. V40_H_H64_G_GTL) >

  10. Go to reconfig. menu and set Bearing type to match the currently installed X-ray Tube. Refer to System Configuration (12HW14.6 SP2.1 or later).

    If NTP server is used, confirm and accept the network setting by Reconfig.

    Reboot is executed during reconfig when bearing type was updated. In this case, skip step 9 “Reboot the system”

  11. Remove the Service Pack CD from Host DVD drive.
  12. Reboot the system:

    {root@hostname]# reboot

  13. Execute “Flash Download” in Utilities tab of CSD menu.
  14. Type “swhwinfo” in command window and confirm that the tube type is same as bearing type in reconfig.

    Example of strings: 13HW31.8_SP2-1-3.V40_NV_V64_G_GTM in case of VCT Liquid Bearing.

    • Ball Bearing : H

    • Liquid Bearing: NV

  15. Perform a Restore System State (Reconfig Info Only) to re-install the options previously removed, then restart the system.
  16. Confirm that the following options has been reinstalled.
    • AdvVesselAnalysis

    • AutoBone

    • CardEP

    • CardIQPlus

    • CTColonoPro

    • CTPerfusion4

    • CTPerfusion4Neuro

    • VolumeViewer

  17. Execute “Restore Runtime Parameters” from Generator tool – JEDI in Diagnostic tab of CSD menu.
  18. Execute “Generator Filament Calibration” in Calibration tab of CSD menu.
  19. Perform sanity scan with the following technique and confirm they completed successfully.

    Scout, Axial and Helical.

4 Service Pack 2.1 Installation (on no SP software)

Procedure

    Notice Image
    • notice
  1. Execute “Save Runtime Parameters” from Generator tool – JEDI in Diagnostic tab of CSD menu.
  2. Execute “Save System State” in Utilities tab of CSD menu.
  3. Uninstall the following options (if installed) using the Install Options menu in the Configuration Tab in Common Service Desktop.
    • AdvVesselAnalysis

    • AutoBone

    • CardEP

    • CardIQPlus

    • CTColonoPro

    • CTPerfusion4

    • CTPerfusion4Neuro

    • VolumeViewer

  4. Open a Unix shell.
  5. If Application software is up, perform the following command in the Unix shell.

    {ctuser@hostname} cleanMon

  6. In the Unix Shell, become root:
    1. {ctuser@hostname} su -
    2. Password: <password>
  7. Insert the Service Pack CD into DVD drive in the host.
  8. Perform the following steps to install the Service Pack CD:
    1. [root@hostname]# patch_install -c
    2. The following messages is displayed.


      These patches from cdrom will be installed:
      13HW31.8_GEHCServicePack1.0
      ------------------------------------------
      I will install update 13HW31.8_GEHCServicePack1.0 from CDROM, is this ok ? [y/n]

                        

    3. Input y to install Service Pack 1.0.
    4. Confirm the “Update 13HW31.8_GEHCServciePack1.0 has been installed successfully” message.
    5. The following messages is displayed.


      These patches from cdrom will be installed:
      13HW31.8_GEHCServicePack2.0
      ------------------------------------------
      I will install update 13HW31.8_GEHCServicePack2.0 from CDROM, is this ok ? [y/n]

                        

    6. Input y to install Service Pack 2.0.

      The following message appears after completion of SP2.0 installation. Ignore it at this timing.


      #################################################################
      # Installation successful.                                      #
      #                                                               #
      # - Execute reconfig in order to set Bearing type (Ball/Liquid) #
      #   after SP2.0 installation completion.                        #
      # - If NTP server is used, confirm and accept the network       #
      #   setting by Reconfig.                                        #
      # - Reinstall the software options and Voxtool protocols by     #
      #   Restore State.                                              #
      #################################################################

                        

    7. Confirm the “Update 13HW31.8_GEHCServciePack2.0 has been installed successfully” message.
    8. The following messages is displayed.


      These patches from cdrom will be installed:
      13HW31.8_GEHCServicePack2.1
      ------------------------------------------
      I will install update 13HW31.8_GEHCServicePack2.1 from CDROM, is this ok ? [y/n]

                        

    9. Input y to install Service Pack 2.1.

      The following message appears after completion of SP2.1 installation.


      #################################################################
      # Installation successful.                                      #
      #                                                               #
      # - Execute reconfig in order to set Bearing type (Ball/Liquid) #
      #   after SP2.1 installation completion.                        #
      # - If NTP server is used, confirm and accept the network       #
      #   setting by Reconfig.                                        #
      #################################################################

                        

    10. Confirm the “Update 12HW14.6_GEHCServciePack2.1 has been installed successfully” message.
  9. Check Software version:
    1. Type: [root@hostname]# swhwinfo ENTER
    2. System should reply with a response as the following:

      13HW31.8_SP2-1-X.<hw info (ex. V40_H_H64_G_GTL) >

  10. Go to reconfig. menu and set Bearing type to match the currently installed X-ray Tube. Refer to System Configuration (12HW14.6 SP2.1 or later).

    If NTP server is used, confirm and accept the network setting by Reconfig.

    Reboot is executed during reconfig when bearing type was updated. In this case, skip step 9 “Reboot the system”.

  11. Remove the Service Pack CD from Host DVD drive.
  12. Reboot the system:

    {root@hostname]# reboot

  13. Execute “Flash Download” in Utilities tab of CSD menu.
  14. Type “swhwinfo” in command window and confirm that the tube type is same as bearing type in reconfig.

    Example of strings: 13HW31.8_SP2-1-3.V40_NV_V64_G_GTM in case of VCT Liquid Bearing.

    • Ball Bearing : H

    • Liquid Bearing: NV

  15. Perform a Restore System State (Reconfig Info Only) to re-install the options previously removed, then restart the system.
  16. Confirm that the following options has been reinstalled.
    • AdvVesselAnalysis

    • AutoBone

    • CardEP

    • CardIQPlus

    • CTColonoPro

    • CTPerfusion4

    • CTPerfusion4Neuro

    • VolumeViewer

  17. Execute “Restore Runtime Parameters” from Generator tool – JEDI in Diagnostic tab of CSD menu.
  18. Execute “Generator Filament Calibration” in Calibration tab of CSD menu.
  19. Perform sanity scan with the following technique and confirm they completed successfully.

    Scout, Axial and Helical

5 Appendix - Service Pack 2.1 Uninstallation

Procedure

    Notice Image
    • notice
  1. Execute “Save Runtime Parameters” from Generator tool – JEDI in Diagnostic tab of CSD menu.
  2. Execute “Save System State” in Utilities tab of CSD menu.
  3. Open a Unix shell and check Exam Split type.

    Type swokinstall –p ENTER

    If the Exam Split option is installed, execute the following.

    Type ls –l ~ctuser/ves/.hesMode ENTER

    If there is .hesMode file, it’s HES mode.

    If not, it’s VES mode.

    This will be used at reinstall option.

  4. Uninstall the following options (if installed) using the Install Options menu in the Configuration Tab in Common Service Desktop.
    • AdvVesselAnalysis

    • AutoBone

    • CardEP

    • CardIQPlus

    • CTColonoPro

    • CTPerfusion4

    • CTPerfusion4Neuro

    • VolumeViewer

  5. In a Unix Shell, type cleanMonEnter to shutdown application.
  6. In a Unix Shell, type su -Enter to be root user.
  7. Type patch_install -u ENTER . Follow the instructions displayed on the screen.

    The following message appears after SP2.1 uninstallation. Execute NTP setting confirmation in reconfig menu as needed.


    #################################################################
    # Uninstallation successful.                                    #
    #                                                               #
    # - If NTP server is used, confirm and accept the network       #
    #   setting by Reconfig.                                        #
    #################################################################

                    

    note:

    It may take 3~4 min., please wait.

  8. Verify all of patches uninstalled with “Successful” message is displayed, then type reboot Enter to reboot the system.
  9. Open a Unix shell, and type swhwinfo ENTER. Confirm that the output does not contain “SP” and will look this:

    13HW31.8.V40_H_HS64_G_GTL.

  10. Perform a Restore System State (Reconfig Info Only) to re-install the options previously removed, then restart the system.
  11. Confirm that the following options has been reinstalled.
    • AdvVesselAnalysis

    • AutoBone

    • CardEP

    • CardIQPlus

    • CTColonoPro

    • CTPerfusion4

    • CTPerfusion4Neuro

    • VolumeViewer

  12. Execute “Flash Download” in Utilities tab of CSD menu.
  13. Execute “Restore Runtime Parameters” from Generator tool – JEDI in Diagnostic tab of CSD menu.
  14. Execute “Generator Filament Calibration” in Calibration tab of CSD menu.
  15. Perform sanity scan with the following technique and confirm they completed successfully.

    Scout, Axial and Helical.

6 Finalization

No finalization steps.