Rotor Diagnostic Test
Overview
The purpose of this test is to drive the rotation inverter in high speed and low speed modes on the tube in order to identify a defective rotation board or a wrong connection between rotation board and tube.
- If desired, you may check any or all of the following:
- PPC kV control board alive and running: DS17-DS24 LEDs are lit successively or a combination of them blink
- Cabling between rotation board and tube checked
- Rotation board alive and running: DS5 LED blinking for rotation V2 (or DS4 LED blinking for rotation V3)
- Rotation board DC supply present: DS7 neon is lit
- Launch the GENERATOR TOOL from the DIAGNOSTICS tab in the Common Service Desktop.
- Launch the Rotor Diagnostic tool.
- Run the test. A rotor sequence will proceed.
- If a problem occurs, check the error log.
1 Finalization
No finalization steps.