JEDI Fan Diagnostics
Overview
The purpose of this procedure is to verify that the generator fans work properly. The procedure consists of removing the right cover and running the test.
- Remove gantry right side cover.
- Turn off the AXIAL ENABLE switch on the Service Switch Panel.
- Safely rotate the gantry so that the inverter and tank fans are visible.
- If desired, you may check that the PPC kV control board is working by looking through the holes on the cover and verifying that S0-S7 LEDs are lit successively or a combination of them blink.
- Launch the GENERATOR TOOL.
- Launch Fans Diagnostic tool.
- Run test. The system will alternate between running the inverter fans and the tank fan.
- Check to see that the fans are working. When a fan is running you can hear it spinning and feel the air movement at the front of the inverter or tank.
1 Finalization
- Change any failed fans.