gesyslog for LB Tube (VCT)
The following gesyslog messages have been added for VCT LB Tube system.
The gesyslog is categorized as “Information” or “Abort”. No. 1 to 3 are Information, No. 4 and 5 are Abort.
1 Tube type
When system started up, Tube type information is posted on gesyslog. The following is example of gesyslog message. If the system uses Hercules Tube, the tube type is “Ball Bearing tube” instead of “Liquid Bearing tube”.
1451322551 0 1 Mon Dec 28 17:09:11 2015 260118447
2
orp orp_subsys
gen_exec.cxx 5030
Mon Dec 28 17:09:12 2015
Host : On-Board Rotating Processor Ermes # : 260118447
Exception Class : Secondary Severity : Pri/Soft
File : gen_exec.cxx Line# : 5030
Function : No System Function Reported
Scan Type : None/Unknown Scan: 0/0/0
Current tube type is Liquid Bearing tube.
2 SGB log
The SGB slope parameter is posted on gesyslog after approx. 10min. later from last rotor start timing. The following is example of gesyslog message.
1451298419 0 1 Mon Dec 28 10:26:59 2015 260100091
2
tgp tgp_subsys
scIsrService.c 198
Mon Dec 28 10:26:59 2015
...skipping...
File : gen_exec.cxx Line# : 9988
Function : No System Function Reported
Scan Type : None/Unknown Scan: 0/0/0
Liquid Bearing tube rotor health log.
SGB Slope: 74 rpm/sec
3 Rotor start without Gantry rotation
LB Tube system requires Gantry rotation for rotor start basically, but some SV Tools require rotor start without Gantry rotation as exception. The following message is posted on gesyslog in this case.
1453979355 1 1 Thu Jan 28 11:09:15 2016 200520025
4
tc03bay environmentMgr
MultiEnvFormMgr.c 286
System enter service mode
...skipping...
Liquid Bearing tube rotor became high speed without the correct rotor management
rule.
A lot of this operation will cause short life of tube.
4 Abort of SV tool execution (Scan request with “rotor must rotate at high”)
When Gantry rotates faster than 2sec./rev., rotor must rotate at high speed in order to prevent fatal rotor damage. SV Tool is aborted by scan request with rotor stop, and the following message is posted on gesyslog.
1454064751 0 1 Fri Jan 29 10:52:31 2016 260100193
2
tgp tgp_subsys
scRxMgr.c 1381
Fri Jan 29 10:52:32 2016
Host : Table Gantry Processor Ermes # : 260100193
Exception Class : Abort Severity : Primary
File : scRxMgr.c Line# : 1381
Function : No System Function Reported
Scan Type : Axial Scan: 50564/1/1
Gantry rotation of faster than 2.0 sec/rot was requested although Liquid Bearing
tube rotor is not rotating.
Rejected the request because it will do fatal damage to the tube.
5 Abort of SV tool execution (Rotor is NOT high speed condition)
When Gantry rotates faster than 2sec./rev., rotor shall rotate at high speed in order to prevent rotor damage. SV Tool is aborted when rotor is NOT high speed, and the following message is posted on gesyslog. This error happens at Gantry Service Balance, Collimator Functional or Axial Functional tools..
1456327617 0 1 Wed Feb 24 15:26:57 2016 260107010
7
tgp tgp_subsys
axial_mgr_events.c 514
Running Gantry Service Balance
Wed Feb 24 15:26:57 2016
Host : Table Gantry Processor Ermes # : 260107010
Exception Class : Secondary Severity : Pri/Soft
File : axial_mgr_events.c Line# : 514
Function : No System Function Reported
Scan Type : None/Unknown Scan: 0/0/0
Shall not rotate Liquid Bearing tube faster than 2 sec/rot
without rotating tube rotor at high speed.
Open command window and execute “rotor_ctrl –start” command
before the procedure execution.