- Topic ID: id_15460937
- Version: 2.0
- Date: Nov 8, 2018 1:36:52 AM
VIG Node Reset
1 VIG Node Process Overview
Use this procedure when replacing a VIG Node, replacing a VDARC Node, or performing a LFC.
VIG Node power off / on is performed with the VDARC communicating (rsh is succesful). If present, use ipmitool command to verify that the ipmi is communicating between the VDARC and the IG Nodes. If a password is required for the VIG Node, select the Enter key, but -P should prevent that field from popping up. Then the vrac_flash_update command is performed. If the vrac_flash_update fails, troubleshoot as needed by verifying ping, rsh, and telnet process as required. Then reset each VIG Node. If a command fails to work become root and retry it.
2 Process
-
If VIG Node is powered ON, turn power OFF at the front panel of each VIG Node.
-
Open a Unix Shell as ctuser on the HOST:
Perform a rsh darc command:
{ctuser@hostname} rsh darc
Last login: Day Month Day Time from oc
[ctuser@darc]$
-
After verifying the DARC Node is up, apply power at the front panel of each VIG Node.
Wait approximately 1.5 minutes to 2 minutes for the VIG Nodes to boot up.
-
Verify that the ipmi is communicating between the VDARC and the IG Nodes
[ctuser@darc ~]$ ipmitool -I lan -H ig1 -P "" sel
SEL Information
Version : 51
Entries : 1155
Free Space : 42432
Percent Used : 30%
Last Add Time : 09/15/2005 13:03:18
Last Del Time : 07/07/2005 03:11:43
Overflow : false
Delete cmd : supported
Partial add cmd : supported
Reserve cmd : supported
Get Alloc Info : supported
# of Alloc Units : 3276
Alloc Unit Size : 20
# Free Units : 2121
Largest Free Blk : 2121
Max Record Size : 8
[ctuser@darc ~]$ ipmitool -I lan -H ig2 -P "" sel
SEL Information
Version : 51
Entries : 1155
Free Space : 42432
Percent Used : 30%
Last Add Time : 09/15/2005 13:03:18
Last Del Time : 07/07/2005 03:11:43
Overflow : false
Delete cmd : supported
Partial add cmd : supported
Reserve cmd : supported
Get Alloc Info : supported
# of Alloc Units : 3276
Alloc Unit Size : 20
# Free Units : 2121
Largest Free Blk : 2121
Max Record Size : 8
[ctuser@darc ~]$ ipmitool -I lan -H ig3 -P "" sel
SEL Information
Version : 51
Entries : 1155
Free Space : 42432
Percent Used : 30%
Last Add Time : 09/15/2005 13:03:18
Last Del Time : 07/07/2005 03:11:43
Overflow : false
Delete cmd : supported
Partial add cmd : supported
Reserve cmd : supported
Get Alloc Info : supported
# of Alloc Units : 3276
Alloc Unit Size : 20
# Free Units : 2121
Largest Free Blk : 2121
Max Record Size : 8
-
Exit the VDARC Node and get back to HOST
[ctuser@darc ~]$ exit
-
Perform the VIG Node vrac flashing process from the HOST
{ctuser@hostname} vrac_flash_update
-
Verify the vrac flashing process was successful.
-
Select the front panel reset button (
) on each VIG Node. -
Verify each VIG Node is successfully able to rsh:
{ctuser@hostname} rsh darc
Last login: Day Month Day Time from oc
[ctuser@darc]$ lhinv | grep Main
(verify Main memory = 6 GB)
[ctuser@darc]$ rsh ig1
[ctuser@ig1]$ lhinv | grep Main
(verify Main memory = 6 GB)
[ctuser@ig1]$ exit
[ctuser@darc]$ rsh ig2
[ctuser@ig2]$ lhinv | grep Main
(verify Main memory = 6 GB)
[ctuser@ig2]$ exit
[ctuser@darc]$ rsh ig3
[ctuser@ig3]$ lhinv | grep Main
(verify Main memory = 6 GB)
[ctuser@ig3]$ exit
-
Perform vrac_diags as follows:
[ctuser@darc]$ su -
Password #bigguy
[root@darc]$ vrac_menu -a (See Note)
Depending on the software version, Cone Beam and Thin Slice diagnostics may not be available from the -a command (above). In this case, perform the following commands with Applications Software down:
vrac_menu -cb –a
vrac_menu –ts –a