- Topic ID: id_15460813
- Version: 2.0
- Date: Nov 8, 2018 1:37:16 AM
VDARC BIOS Telnet Process
If the SCSI BIOS "extended int 13 support for drives > 1GB" setting is disabled on either host controller, then gre-raid will fail to create/start the scan disk array (by design).
SCSI CONTROLLER SETTINGS for A and B “Sync Transfer Rate should all be at 320 (not 160).
Description:
Ideally, this one-time procedure on the VDARC Node is done before the LFC. Use Serial Over LAN (SOL) to reboot VCT DARC Node and access the DARC SCSI setup.
-
With Application Software DOWN:
-
Open a Unix Shell.
-
telnet localhost 623 (cli service may need to be restarted if this fails)
-
Server: darc
-
Username: (press the Enter/Return key)
-
Password: (press the Enter/Return key)
-
At the dpccli prompt: reset -f -c
Press OK.
-
-
Press Ctrl+A (when message appears) to configure the SCSI BIOS (appears shortly after system BIOS display)
-
AIC-7902 A at slot ## xx:xx:xx
-
Press Enter to select AIC-7902 A at slot ##.
-
Press Enter to go to the Configure/View SCSI Controller Settings screen.
-
Use down-arrow key to select SCSI Device Configuration.
-
Verify the Sync Transfer Rate (MB/Sec) for all is set at 320.
If not:
-
Arrow to the value in need of change and press Enter.
-
Use up arrow key to 320 and press Enter then arrow over and select Enter for next change.
-
Verify all (0 through 15) are set to 320.
-
Press Esc key to leave the Configuration screen for AIC-7902 A.
-
Press Esc Key to leave this screen and return to AIC-7902 A at slot ## Screen.
-
-
Use down-arrow key to select Advanced Configuration and press Enter.
-
Use down-arrow key to select Extended Int 13 Translation for DOS Drives > 1 GByte.
-
If setting is disabled, press Enter and select/enter Enabled.
-
Press ESC key twice to go up two levels: (AIC-7902A Configuration screen then Save Changes Made?).
-
Select Yes to save changes.
-
Press Esc key to get out of the options screen: Configure/View SCSI Controller Settings.
-
-
AIC-7902 B at slot ## xx:xx:xx
-
Press Enter to select AIC-7902 B at slot ##.
-
Press Enter to go to the Configure/View SCSI Controller Settings.
-
Use down-arrow key to select SCSI Device Configuration.
-
Verify the Sync Transfer Rate (MB/Sec) for all is set at 320.
If not:
-
Arrow to the value in need of change and press Enter.
-
Use up arrow key to 320 and press Enter then arrow over and select Enter for next change.
-
Verify all (0 through 15) are set to 320.
-
Press Esc key to leave the Configuration screen for AIC-7902 B.
-
Press Esc key to leave this screen and return to AIC-7902 B at slot ## Screen.
-
-
Use down-arrow key to select Advanced Configuration and press Enter.
-
Use down-arrow key to select Extended Int 13 Translation for DOS Drives > 1 GByte.
-
If setting is disabled, press Enter and select/enter Enabled.
-
Press ESC key twice to go up two levels. It displays - (AIC-7902A Configuration screen then Save Changes Made?)
-
Select Yes to save changes.
-
Press Esc key to get out of the options screen: Configure/View SCSI Controller Settings.
-
-
Press Esc key to leave this screen and return to AIC A/B Selection Screen.
-
Press Esc key to leave the EXIT UTILITY. Use Arrow Key to select Yes and press Enter key.