• Topic ID: id_15460622
  • Version: 2.0
  • Date: Nov 8, 2018 1:36:23 AM

Image Quality Test

1 Preparation

  • All table mechanical alignment procedures are completed.

  • The table perpendicular alignment test passed.

  • The table tilt alignment test passed.

  • Table anchors are in place and within specification.

2 Procedures

Notice Image
  • notice
  • Run ALL Image Series Tests in Auto Mode.
  1. Check that the table cradle is level in all directions. Correct, if necessary.

  2. Check the scan window for proper installation

  3. From the operator’s desktop, select SCANNER UTILITY.

  4. Complete COLLIMATOR CALIBRATION and FAST CALIBRATION.

  5. Complete CENTER PHANTOM procedure.

  6. Locate the white service CD that ships with the system, located in the tech pub tray of the lean cart.

  7. Load the CD into the laptop drive. Look under the Functional Checks tab located on the left hand side. Run the QA procedure.

  8. Record all pass/fail information on the GE e4879 form.