- Topic ID: id_15460622
- Version: 2.0
- Date: Nov 8, 2018 1:36:23 AM
Image Quality Test
1 Preparation
-
All table mechanical alignment procedures are completed.
-
The table perpendicular alignment test passed.
-
The table tilt alignment test passed.
-
Table anchors are in place and within specification.
2 Procedures
![]() |
|
-
Check that the table cradle is level in all directions. Correct, if necessary.
-
Check the scan window for proper installation
-
From the operator’s desktop, select SCANNER UTILITY.
-
Complete COLLIMATOR CALIBRATION and FAST CALIBRATION.
-
Complete CENTER PHANTOM procedure.
-
Locate the white service CD that ships with the system, located in the tech pub tray of the lean cart.
-
Load the CD into the laptop drive. Look under the Functional Checks tab located on the left hand side. Run the QA procedure.
-
Record all pass/fail information on the GE e4879 form.