General Gantry Inspection
1 Listen for Unusual Noises
![]() |
|
-
Use the gantry service switch panel to rotate the gantry at the slow position.
-
Listen for any unusual gantry sounds, such as:
-
Bearing noise
-
Rubbing components
-
Rotating components touching covers
-
Loose components
-
Any other unusual sounds
-
-
Repeat the procedure at the fast position.
2 Verify Display Lights and E-Stop Buttons
-
Using the gantry control push-button, advance the cradle about 2 ft. (610 mm) from the home position.
-
Press one of the emergency stop switches on the gantry controls.
-
Depress one of the tilt buttons. Verify that tilt is disabled.
-
Depress one of the table elevation buttons. Verify that table elevation is disabled.
-
Depress one of the cradle drive buttons. Verify that cradle drive is disabled.
-
Verify that the cradle clutch is released by manually moving the cradle to the home position.
-
Clear the hazard by pressing the reset button on the gantry control panel.
-
Verify that tilt, elevation, and cradle drive now work.
-
Press the remaining E-stops (4 on the gantry, 1 on the PDU, 1 on the SCIM). Verify that power is removed each time by observing the gantry display.
note:If any of the E-stops do not pass, you need to schedule a time to replace the E-stop button. (For replacement procedures, see Service Outlet and E-Stop Replacement.)
3 Test Patient Tilt Sensors
There are two Patient Tilt Sensors: one in front and one in back. Use the following steps to test both sensors.
-
From the front cover control panel, press and hold the gantry Forward Tilt button.
-
Press on the Patient Tilt Sensor while still tilting the gantry.
-
Verify the gantry tilt stops.
-
Release both the Patient Tilt Sensor and the Tilt button.
-
Press the Forward Tilt button and verify that the gantry tilts forward.
-
From the rear gantry control, press and hold the Backward Tilt button.
-
Press on the rear Patient Tilt Sensor while still tilting the gantry.
-
Verify that the tilt stops.
-
Release both the Patient Tilt Sensor and the Tilt button.
-
Press the Backward Tilt button and verify that the gantry tilts backward.
-
Return the gantry tilt to zero by pressing and holding the Forward Tilt and Backward Tilt buttons at the same time.