- Topic ID: id_18718307
- Version: 2.1
- Date: Aug 18, 2020 3:42:43 AM
Table Elevation Encoder Replacement
Prerequisites
Overview
Procedure Effectivity:
Procedure
- Raise the table to maximum height.
- Remove gantry covers as required.
Refer to
warning- notice
- Remove power from table by turning off 120VAC, Axial Drive and HVDC switches on Service Switch Panel.
- Remove Base Covers and left Side Panels.
- Remove Encoder Assembly, as described in Elevation Encoder Assembly Replacement.
- To loosen the encoder, turn the three screws holding it to the mounting block CCW, 180°.
- Loosen the screw on the encoder side of the flexible coupling.
- Remove defective Elevation Encoder.
- Install the new Elevation Encoder on the mounting bracket using three screws.
- Replace the Encoder Assembly as described in Elevation Encoder Assembly Replacement.
![]() |
Finalization
-

- notice
- Power up the Table from the Service Switch Panel.
- Follow the procedure on Table Characterization to Characterize the Table Elevation.
- Using the service switches on the MTCB, move the Table up and down, and the Cradle in and out.
- Use the gantry control keys to move the Table up and down, and the Cradle in and out.
- When the Table functions properly, reinstall the Table Base Covers, the Side Panels and the Gantry Cover(s).