- Topic ID: id_16158122
- Version: 1.0
- Date: Jul 7, 2018 4:43:14 PM
TGPU Replacement
Prerequisites
Overview
1 Removing The TGPu
Procedure
caution- Remove all gantry covers except for the front cover.
- notice
- Turn off all three (3) switches (Axial Drive, HVDC, and 120VAC) on the Service Switch Panel.
- Remove the 12 screws in the TGPU covers.note:
Front and rear access is required for this procedure.
- Remove the cover.
- Disconnect all cabling into the TGPU.note:
If the cables are not labeled, label them as they are removed.
- Remove the eight (8) screws attaching the TGPU board.
- Remove the defective TGPU.
![]() |
2 Install TGPu
Procedure
- Put the new TGPu board in place.
- Secure the board by replacing the eight (8) screws.
- Carefully install all of the removed cables, making sure that no pins are bent during engagement.
- Replace the cover.
- Replace the twelve (12) screws in the TGPU cover.note:
Front and side access may be required.
- Execute Flash download query and update. TGPu firmware older than OC
may fail. In this case, execute the following:note:
Before powering ON the 120VAC gantry power, open a shell and type the following to perform the Flash download procedure so that TGPu firmware is downloaded from the OC.
cd /usr/g/fw Enter
touch tgpFtp Enter
- Turn ON all three (3) switches (Axial Drive, HVDC, and 120VAC) on the Service Switch Panel.
3 Finalization
Procedure
- Reinstall covers.note:
Remove LOTO.
- Perform System Scanning Test.