Streak Test
Prerequisites
Overview
This document provides the necessary steps to test for streak artifacts.
1 Phantom Setup
Procedure
- Install all gantry covers and scan window.
- Move table to home longitudinal position.
- Attach phantom holder to gantry end of cradle.
- Mount and center phantom to phantom holder.
- Place bubble level on top of phantom, positioned along the Z-axis. Adjust phantom holder back knob until phantom is level front to back.
- Turn on laser alignment lights.
- Move table into gantry until axial internal alignment lights are lined up with phantom's axial reference mark.
- On the console, select SCANNER UTILITIES from the bottom of the Scan Monitor.
- Select CENTER PHANTOM from the Scanner Utilities window.
- Select CONFIRM in the Phantom Centering window.
- Press START] on SCIM when it flashes. A Phantom Centering scan is performed.
- Adjust phantom position if results are out of specification
range.note:
The phantom center specification is ± 0.5mm.
- Repeat steps [10] – [12] until phantom is centered.
- Once phantom is centered, press DONE in the Phantom Centering window.
- Press QUIT in the Scanner Utilities window.
- Set Internal Landmark.
2 Acquire Scans
Procedure
- Click NEW PATIENT from the bottom of the Scan Monitor.
- Type the following entries in the listed Patient Information fields:
- Patient ID: Service
- Patient Name: Streak Test
- From the Protocol Selection display, click on the Anatomical Selector Service tab.
- On the Service Protocol Selection window,
select IMAGE QUALITY, 44.1 STREAK
TEST.note:
Make sure the prescription is a single group axial protocol.
- Select CONFIRM. An Attention message box is displayed.
- Click CONTINUE in the Attention message box.
- Press START on the SCIM when it flashes. Scans are performed.
- When all the scans are completed, click END EXAM.
- If a PPS Response window is displayed, click COMPLETE.
3 Image Review
Procedure
- From the IMAGEWORKS desktop List/Select browser, select the scanned examination.
- Select VIEWER and review all images from the scanned examination for streak artifacts.
- If any images depict streaks, perform the steps in Section 4.4 on those images.
4 Streak Analysis
Procedure
- Select SERVICE desktop and move the Service Desktop window to the left monitor.
- Select IMAGE QUALITY from the Service Desktop.
- Select IMAGE ANALYSIS and move IMAGE ANALYSIS window to the left monitor.
- From the List/Select browser on the right monitor, select the image to be analyzed.
- Click theMANUAL radio button on the Image Analysis window.
- Click the MANUAL selection button, and select STREAK.
- Click ACCEPT. The image is displayed in the Image Analysis window with a Streak ROI.
- Follow instructions, in Image Analysis window lower right pane, to size Streak ROI and analyze.
- From the results, if the Measured Streak Factor is out of the specification range of +/-4.0, perform appropriate corrective action.
- When all tests are completed, click QUIT.
5 Finalization
No finalization steps.