ORP Board Replacement
Prerequisites
Overview
Procedure
- Remove gantry covers as required.
Refer to
- notice
- Turn off all three switches (Axial Drive, HVDC, 120VAC) on the Service Control Panel.
- Position ORP at the 2 o'clock position.
- Put on the grounding wrist strap.
- Remove cabling from front of the ORP Assembly.
- Remove the two (2) large screws to remove the ORP Assembly.
- Remove the two (2) cover screws and remove cover.
- Remove seven (7) screws connecting the ORP board to the ORP Assembly.
- Remove the defective ORP board.
- Install the new ORP board in reverse order.
Set the torque on the 6mm cap screws to 2.9 N-m (25.66 in-lbs)
![]() |
Finalization
- Remove LOTO and Power-Up system.
- On the console, run Flash Download.
note:
Before powering ON the 120VAC Gantry power, open a shell and type the following to perform the FLASH download procedure so that ORP firmware is downloaded from the OC.
cd /usr/g/fw [Enter]
touch orpFtp [Enter]
- Perform System Scanning Test (located in folder).