• Topic ID: id_18718304
  • Version: 2.0
  • Date: Feb 4, 2020 1:40:51 AM

MTCB Cover Replacement

Prerequisites

Overview

Procedure

  1. Raise the table to maximum height.
  2. Remove gantry covers as required.

    Refer to Parts Replacement > Gantry > Enclosure > (Cover Removal Procedure).

  3. Notice Image
    • notice
    • Always turn OFF the HVDC before the 120 VAC. Turning OFF 120 VAC power before HVDC power can result in equipment damage.
  4. Remove power from table by turning off 120VAC, Axial Drive and HVDC switches on Service Switch Panel.
  5. Remove right Table Base Cover.
  6. Remove the two (2) screws that hold the MTCB Cover in place.
  7. Insert new MTCB Cover, reassemble and power-on Table.

Finalization

  1. Confirm that the table is functioning properly.