MOD Peripheral Tower Installation

Prerequisites

Overview

This procedure describes and illustrates the steps necessary to install the Option MOD Peripheral Tower.

Figure 1. MOD Peripheral Tower

1 Power-Off (Shut Down) the Console

Procedure

  1. Select one of the following methods to power off the Operator Console:
    • If applications are running, click the Shut Down icon and select Shut Down.

    • If applications are down, open a Unix Shell using the Toolchest. Type: {ctuser@hostname} halt. Press Enter.

    The Operator Console monitor will display a ‘System Halted’ message when it is acceptable to power off the Operator Console.

  2. Power OFF the Operator Console at the front panel switch. (See Figure 2.)

    Figure 2. Console Power Switch

2 Install Option MOD Peripheral Tower

Procedure

  1. Install the Option MOD Peripheral Tower chassis on the Operator Console.
  2. Mount the power cord to the rear of the MOD Peripheral Tower chassis. Verify that the power switch on the MOD Peripheral Tower Power Supply is turned to the ON position.
  3. Make sure the rear SCSI cable connections.

    Figure 3. MOD Peripheral Tower Connections

3 Power-On the Operator Console

Procedure

  1. Power ON the Operator Console at the console front panel switch.
  2. Visually verify the MOD Peripheral Tower front panel Power LED is illuminated.

    Figure 4. MOD Peripheral Tower Power LED

4 Verify Install MOD Peripheral Tower Operation

Procedure

  1. Visually verify the Peripheral Tower front panel Power LED is illuminated and that each of the MOD Drive Access LEDs illuminates when media is inserted and the tray is closed.
  2. No other verifications required.

5 Attach Label on MOD Tower

The label Part Number is 5346627. It is shipped with TIO Console, not in the Option kit. So please find the label from the TIO Console package. Paste the label behind the MOD Tower, proper location refer to Figure 5

Figure 5. MOD Tower Label

Only trained GE personnel should be install and service the power connection of the MOD Tower.