Home Latch Assembly Replacement
Prerequisites
Overview
Procedure Effectivity:
Procedure
- Raise the table to maximum height.
- Remove right Gantry Side Cover.
warning- notice
- Remove power from table by turning off 120VAC, Axial Drive and HVDC switches on Service Switch Panel.
- Remove the right Table Side Rail Covers and Cradle Drive Cover, to access the Cal Pin.
- Install and tighten the Cal pin to about one-half inch from the home position, to hold the carriage in place.
- Remove the Cradle Assembly and the right Side Rail Cover.
- Disconnect the harness at the terminal strip.
- Remove the four screws that fasten the Latch Assembly to the z-channel.
- Remove the defective Latch assembly.
- Install the replacement Latch Assembly, but do not tighten the four screws.
- Adjust the position of the entire Latch Assembly in its slots:
- Adjust the Latch assembly until the distance between the carriage latch block and the forward edge of the latch bar opening equals 0.050 ±0.005 inches.
- Maintain this distance while you tighten the four screws.
- Adjust the set screw in the latch clevis block, until the outer edge of the latch bar overlaps the outer edge of the carriage latch block by 0.050 ±0.00 inches.
- Maintain this distance while you tighten the jam nut.
- Make sure the solenoid plunger bottoms out:
- Adjust the position of the solenoid bracket, until the clearance between the outer edge of the latch bar and the outer edge of the carriage latch block equals 0.050 ±0.005 inches.
- Maintain this distance while you tighten the two screws.
- Adjust the position of the spring bracket, until the spring has 0.125
inches pre-load, when the latch bar rests against the set screw.
Maintain this distance while you tighten the two screws.
- Install the Cradle.
- Return the Cal pin to its storage position.
- Return power back to the table and replace table and gantry covers.
![]() |
Finalization
-

- notice
- Power up the Table from the Service Switch Panel.
- Using the service switches on the MTCB, move the Table up and down, and the Cradle in and out.
- Use the gantry control keys to move the Table up and down, and the Cradle in and out.
- When the Table functions properly, reinstall the Table Base Covers, the Side Panels and the Gantry Side Cover.