E-Stop Button Replacement
Prerequisites
Overview
Procedure
- Remove gantry covers as required.
Refer to
- notice
- Turn off all three (3) switches (Axial Drive, HVDC, 120VAC) on the Service Switch Panel.
- Loosen two (2) screws to remove the two (2) cables.
- Loosen the four (4) hex screws that fasten the emergency stop button to the cover.
- Replace the emergency stop button and reconnect cables.
- notice
- Reassemble gantry.
![]() |
![]() |
Finalization
- Confirm that all E-stop buttons are operational.