Collimator Calibration
Prerequisites
This document provides necessary steps to detailed calibrate the system.
Procedure
- Move table to longitudinal home position.
- Click SCANNER UTILITIES.
- Click COLLIMATOR CALIBRATION.
- Click OK in Attention window.
- Press START on SCIM when it flashes.
- Click OK in Attention window.
- Click QUIT in Scanner Utilities window.
Finalization
- Detailed Calibration
- GSI Detailed Calibration
- Fast Calibration
- Save System State