- Topic ID: id_18718126
- Version: 1.1
- Date: Aug 18, 2020 3:10:26 AM
5V Power Supply Replacement
Prerequisites
Overview
Procedure Effectivity:
Procedure
- Raise the table to maximum height.
- Remove right Gantry Side Cover.
warning- notice
- Remove power from table by turning off 120VAC, Axial Drive and HVDC switches on Service Switch Panel.
- Remove the right Table Base Cover.
- Remove the MTCB Cover to ease access to the 5V Power Supply.note:
Pay attention to the locations of the wires on the defective power supply, before you move them. Restore the wires to their original configuration on the replacement power supply.
- Remove wires connected to the 5V Power Supply.
- Remove the two screws connecting the Power Supply Mounting Bracket to the base of the table.
- Remove the old power supply and the Mounting Bracket.
- Remove the three screws from the Mounting Bracket to release the old Power Supply.
- Screw the new 5V Power Supply into the Mounting Bracket.
- Screw the new Power Supply and Mounting Bracket into the base of the Table.
- Reconnect all wires back to the 5V Power Supply.
- Return the MTCB Cover to its original position.
![]() |
Finalization
-

- notice
- Power up the Table from the Service Switch Panel.
- Using the service switches on the MTCB, move the Table up and down, and the Cradle in and out.
- Use the gantry control keys to move the Table up and down, and the Cradle in and out.
- When the Table functions properly, reinstall the Table Base Covers, the Side Panels and the Gantry Side Cover.