- Topic ID: id_18718122
- Version: 1.0
- Date: Sep 20, 2018 2:05:34 PM
24V Power Supply Replacement
Prerequisites
Procedure
- Raise the table to maximum height.
- Remove right Gantry Side Cover.
warning- notice
- Remove power from table by turning off 120VAC, Axial Drive and HVDC switches on Service Switch Panel.
- Remove right and left Table Base Covers.note:
Pay attention to the locations of the wires on the defective power supply, before you move them. Restore the wires to their original configuration on the replacement power supply.
- Remove wires connected to the 24V Power Supply.
- Remove the two screws connecting the 24V Power Supply Mounting Bracket to the base of the table.
- Remove the old Power Supply and the Mounting Bracket.
- Remove the three screws from the Mounting Bracket to release the old Power Supply.
- Insert the new Power Supply into the Mounting Bracket.
- Screw the new Power Supply and Mounting Bracket back into the base of the Table.
- Reconnect all wires back to the Power Supply.
![]() |
Finalization
-
Power up the Table from the Service Switch Panel.
- notice
- Using the service switches on the MTCB, move the Table up and down, and the Cradle in and out.
- Use the gantry control keys to move the Table up and down, and the Cradle in and out.
- When the Table functions properly, reinstall the Table Base Covers, the Side Panels and the Gantry Side Cover.