• Topic ID: id_18718528
  • Version: 1.0
  • Date: Sep 20, 2018 2:02:26 PM

Verify Disk Array Drives

Open a Unix Shell and type the following, to verify Disk Array drives are present:

{ctuser@ hostname }   rsh  darc

Last login: Tue Oct 7 14:58:34 from OC

{ctuser@darc}   cat  /proc/scsi/scsi/

note:

The output (shown below) may differ, due to various manufacturers of approved production released SCSI disk drives. However, two (2) 36 GByte drives must be present for a good Disk Array.

An example output is shown below.

Attached devices:

Host: scsi0 Channel: 00 Id: 00 Lun: 00

Vendor: MITSUMI Model: CD-ROM SR244W Rev: T01A

Type: CD-ROM ANSI SCSI revision: 02

Host: scsi1 Channel: 00 Id: 00 Lun: 00

Vendor: SEAGATE Model: ST3 36 753LW Rev: 0005

Type: Direct-Access ANSI SCSI revision: 03

Host: scsi1 Channel: 00 Id: 01 Lun: 00

Vendor: SEAGATE Model: ST3 36 753LW Rev: 0005

Type: Direct-Access ANSI SCSI revision: 03