- Topic ID: id_18718528
- Version: 1.0
- Date: Sep 20, 2018 2:02:26 PM
Verify Disk Array Drives
Open a Unix Shell and type the following, to verify Disk Array drives are present:
{ctuser@ hostname } rsh darc
Last login: Tue Oct 7 14:58:34 from OC
{ctuser@darc} cat /proc/scsi/scsi/
The output (shown below) may differ, due to various manufacturers of approved production released SCSI disk drives. However, two (2) 36 GByte drives must be present for a good Disk Array.
An example output is shown below.
Attached devices:
Host: scsi0 Channel: 00 Id: 00 Lun: 00
Vendor: MITSUMI Model: CD-ROM SR244W Rev: T01A
Type: CD-ROM ANSI SCSI revision: 02
Host: scsi1 Channel: 00 Id: 00 Lun: 00
Vendor: SEAGATE Model: ST3 36 753LW Rev: 0005
Type: Direct-Access ANSI SCSI revision: 03
Host: scsi1 Channel: 00 Id: 01 Lun: 00
Vendor: SEAGATE Model: ST3 36 753LW Rev: 0005
Type: Direct-Access ANSI SCSI revision: 03