- Topic ID: id_16157951
- Version: 1.0
- Date: Jul 7, 2018 4:33:36 PM
Service Pack 1.7 for 09BW35.11 Installation (only for service pack installation)
1 Overview
This procedure applies to the systems with 09BW35.11 already installed.
2 Check System Configuration
![]() |
|
![]() |
|
Check system configuration by one of the following methods:
-
Open a Unix Shell and type one of the following:
check_config
- OR -
swhwinfo -all
-
Look at the Common Service Desktop Home Page.
3 Procedure
-
Save System States.
-
Remove all options on system
note:some options might not function well if options are not removed this time and restored back after SP1.7 installed.
-
Open a Unix Shell.
-
If Application software is up, perform the following command
{ctuser@ hostname } cleanMon
-
In the Unix Shell and become root:
-
{ctuser@ hostname } su -
-
Password: <password>
-
-
Check whether Service Pack is installed or not. Type [root@hostname]#swhwinfo
-
If SP1-2-1 or SP1-4-1 included in the output, uninstall it using the instruction below first
[root@ hostname ] #patch_install -u, the run step 7.
-
If SP1-5-1 included or no information like SP X-X-X included in the output, move to step 7.
-
-
Insert the Service Pack CD into DVD drive in the host.
-
Perform the following steps to install SP1.3, SP1.5, SP1.7 orderly:
-
[root@ hostname ] #patch_install -c
-
Input y to install patch.
-
-
Type: [root@hostname]#reboot
-
After the application is up, restore system states
-
Reboot system, after the system is up, perform Flash Download from Common Service Desktop.