- Topic ID: id_18718527
- Version: 1.0
- Date: Sep 20, 2018 2:02:34 PM
Reconfig Note
If the number of IG Nodes present is incorrect, then perform the reconfig process as root on the Host with Application Software down.
{ctuser@ hostname } su -
Password: #bigguy
[ctuser@root] reconfig
Select the Hardware Tab
Correct for the number of IG Nodes actually present in the Operator Console.
Select Accept Tab
The Operator Console will reboot