• Topic ID: id_17423310
  • Version: 3.0
  • Date: Apr 22, 2019 12:56:03 AM

Real-Time Analog Signals (LS16, Ultra, Plus)

The JEDI60DC provides real-time analog kV and mA signals to the DAS for inclusion in the image data stream. These are negative polarity differential pair signals with 1k ohm source impedance. Scaling for the kV signal is 0 to -10Vdc for 0 to 200kV and 0 to -10Vdc for 0 to 1000mA.