• Topic ID: id_11038958
  • Version: 2.0
  • Date: Jan 30, 2019 9:42:15 PM

Example Config/INFO Output

This section contains an example system hardware configuration INFO file output that is displayed when you type the following in a shell: cat /usr/g/config/INFOEnter.

For some parameters, the values for your system may differ from those in the example.


setenv SERVICE_ID GE_SERVICE
setenv HOSPITAL_NAME                        ________________
setenv MENUTZ "America"                     ________________
setenv SUNTZ "US/Central"                   ________________
setenv TZ "CST6CDT US"                      ________________
setenv TZTEXT "Chicago"                     ________________
setenv REGENERATE_DATABASE no
setenv NEXT_MOD_NUMBER 1
setenv NEXT_DIAGNOSTIC_EXAM_NUMBER 50000
setenv GATEWAY_HOSTNAME                     ________________
setenv GATEWAY_IP                           ________________
setenv GATEWAY_NETMASK                      ________________
setenv GATEWAY_BROADCAST                    ________________
setenv GATEWAY_DEFAULT                      ________________
setenv TUBETYPE 15 (Performix) / 16 (Hercules) / 17 (OEM) ______________
setenv PDUTYPE CT_Compact / NG PDU          ________________
setenv DASMTYPE 2
setenv KV_VALUES "80 100 120 140"           ________________
setenv DEFAULTARCHIVE DICOM
setenv ADVANTAGERW FALSE
setenv CTIDATETIMEFORMAT 4
setenv TIMESEPAR /
setenv WEIGHT lb
setenv DOCTITLE "radiologist"
setenv CAMERATYPE ""
setenv SUITEID                              ________________
setenv FILTERTYPE 2
setenv CAMERA no
setenv DEFFORMAT 12
setenv DOGREYSCALE yes
setenv DOSLIDES no
setenv DOZOOM no
setenv FILMINTERP sharp
setenv PRINTER_HOSTNAME                        ________________
setenv PRINTER_IPADDRESS                       ________________
setenv PRINTER no
setenv O2_PRESENT n
setenv GANTRY_TYPE h2 / hp/ wb                 ________________
setenv LANGUAGE en_US
setenv DOSE y
setenv NETWORK yes
setenv USENIS yes
setenv CHGDARCSUBNET yes
setenv DOMAINNAME                              ________________
setenv NIS_SERVER                              ________________
setenv DARC_SUBNET                             ________________
setenv NEXT_PATIENT_EXAM_NUMBER 1
setenv MOBILESYSTEM no                         ________________
setenv STREAK y
setenv INROOMSCAN off
setenv AAA n
setenv AUTOMA_INDEX_TABLE_TYPE type2           ________________
setenv RECONHARDWARE GRE / No O2 Pegasus       ________________
setenv DASTYPE SDAS or MDAS 4 / 8/ 16          ________________
setenv DETECTORTYPE Warp3
setenv REFRESH_RATE 76

setenv yp s
setenv ypmaster ${SUITEID}_OC0
setenv HOSTNAME SBC0
setenv ALIAS ${SUITEID}_SBC0
setenv UNSYSID $GATEWAY_HOSTNAME
setenv SERVID $GATEWAY_HOSTNAME
setenv tubeType $TUBETYPE
setenv ACRNEMA xxxx
setenv REGEN no

setenv HRCTTITLE                               ________________
setenv HRPORTNUM                               ________________
setenv HRAETITLE                               ________________
setenv HRIPADDR                                ________________
{ctuser@hosatname}