• Topic ID: task_ejl_vmd_zhb
  • Version: 5.0
  • Date: May 9, 2022 11:43:44 PM

Z8G4 Card Replacement

Prerequisites

Overview

This procedure shall be followed when replacing the Card in a Z8G4 computer.

Figure 1. Card

1 Preparation

Procedure

  1. Shutdown system. Select one of the following methods to Power OFF the Console:
    • If Applications are up, click on the [Shut Down] button on desktop display and select [Shutdown].

    • If Applications are down, open a Terminal Window. Type: halt , then press ENTER.

    • When halt command has finished, power Off the console at the front panel switch.

  2. Apply LOTO. See Equipment Service - Lockout-Tagout-PPE procedure.

2 Card Removal

Procedure

  1. Open and remove the module small side cover

    Figure 2. Remove module small side cover

  2. Locate Card by referencing the Illustration below.

    Figure 3. Z8G4 Component Location

  3. Pull out the old card from the host computer (slot 1).

3 Card Install

Procedure

  1. Reinstall the card.
  2. Reinstall all removed components, and reconnect any cables that have been disconnected.
  3. Remove LOTO from console.

4 Finalization

Procedure

  1. Reinstall module small side cover
  2. Reinstall the host computer left side panel.
  3. Confirm Host Computer powers up when console power is turned on.
  4. Perform the Functional Checks → System Scanning Test instructions from the procedure list.